Publikationsansicht

A NANO INTERDIGITATED ELECTRODES ARRAY ON POLYMER FOR DISPOSABLE IMPEDIMETRIC BIOSENSORS (2008)

Abstract
This paper presents an integrated nano interdigitated electrodes array (IDA) on polymer substrate with microfluidic system. It can be used as a miniaturized, sensitive, and easy-to-use impedimetric sensor for genomics, proteomics, and cellular analysis. A gold (Au) nano IDA has been successfully patterned on polymer (Cyclic olefin copolymer, COC) substrate, which can be widely used for disposable lab-on-a-chip applications. The fabricated device has been characterized in KCl salt solution using electrochemical impedimetric spectroscopy (EIS). The preliminary result of monitoring protein binding is demonstrated as well, which insures the potential of this device to be used for protein immunosensors.

Details der Publikation
Download http://citeseerx.ist.psu.edu/viewdoc/summary?doi=?doi=10.1.1.125.6281
Quelle http://www.ececs.uc.edu/~zouz/Hilton Head 2006.pdf
Mitarbeiter CiteSeerX
Archiv CiteSeerX - Scientific Literature Digital Library and Search Engine (United States)
Typ text
Sprache Englisch